Máy đếm hạt bụi Lazer
Model: MET ONE 3400
Hãng sản xuất: Beckman Coulter – Mỹ
Standard Size Channels
|
3413 and 3423 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm
|
3415 and 3425 0.5, 1.0, 2.0, 3.0, 5.0, 10.0 or 0.5, 1.0, 2.0, 3.0, 5.0, 25.0* μm
| |
3445 0.5, 1.0, 2.0, 3.0, 5.0, 10.0um
| |
Flow Rate
|
3413 and 3415 28.3 LPM (1.0 CFM) ± 5% (default factory setting)
|
3423 and 3425 50 LPM (1.77 CFM) ± 5% (default factory setting)
| |
3445 100LPM (3.53 CFM) ± 5% (default factory setting)
| |
Zero Count
|
Conforms to JIS B9921. 1 count or less in 5 minutes, 95% confidence level
|
Coincidence Loss
|
28.3 LPM (1.0 CFM) 5% at 14,126,000 particles/m3 (400,000 particles/ft3) 50 LPM (1.77 CFM) 5% at 4,000,000 particles/m3 (113,266 particles/ft3) 100 LPM (3.53 CFM) 10% at 10,000,000 particles/m3 (283,166 particles/ft3)
|
Counting Efficiency
|
3413 and 3423 50% ± 20% for 0.3um (100% ± 10% at 1.5 times the minimum sensitivity), fully complies with ISO 21501-4
|
3415, 3425 and 3445 50% ± 20% for 0.5um (100% ± 10% at 1.5 times the minimum sensitivity), fully complies with ISO 21501-4
| |
Light Source
|
Long Life Laser™ diode with 10-year Mean Time To Failure (MTTF)
|
Pump Type
|
Air vacuum, rated for continuous use
|
Count Display
|
Color 1/4 VGA TFT Touch Screen
|
Interface
|
Windows CE®-based
|
Sample/Hold/Delay Times
|
1 second to 23 hrs 59 minutes 59 seconds
|
Count Alarms
|
1 to 9,999,999 counts
|
Count Cycles
|
Up to 100 while in automatic mode
|
Location Labels
|
0 to 999, appears on printout
|
Data Storage
|
5,000 samples, scrollable on Historical Data review screen
|
Communication Interfaces
|
RS232, RS485, 802.11 b/g Wireless, 802.3 Ethernet, USB Client (Version 1.1), USB Host (Version 1.1)
|
Communication Protocol
|
Modbus TCP, Modbus RTU, Serial FX
|
Battery Type
|
Lithium ion smart battery; rechargeable, ejectable and hot-swappable
|
Operating Time (Battery)
|
28.3 LPM (1.0 CFM) 6 hours** 50 LPM (1.77 CFM) 6 hours** 100 LPM (3.53 CFM) Up to 3 hours**
|
Battery Recharge Time
|
6.75 hours minimum, 10 hours maximum
|
Power
|
24 VDC 3.2A with 100~240 VAC 50/60 Hz adapter (included in ship kit)
|
Size, Dimensions
|
31.8 W x 25.4 D x 20.3 H cm (12.5 x 10 x 8 inches)
|
Weight (without battery)
|
3413 and 3415 7.55 kg (16.6 lbs) 3423 and 3425 8.33 kg (18.3 lbs) 3445 8.65 kg (19.0 lbs)
|
Enclosure Material
|
Stainless Steel
|
Environment
|
Operating 0 to 40° C (32 to 104° F), 10 to 90% relative humidity, non-condensing Storage -40 to 50° C (-40 to 122° F), 0 to 98% relative humidity, non-condensing
|
* Channel selections can be selected at time of order in a range from 0.3 μ to 25 μ. However 0.3 μ and 25 μ cannot be configured together.
| |
** Battery life is estimated with these conditions: The 1.77 cfm unit sampling for 20 minutes (1 m3) sample, print record, a 5-minute hold time (simulating move to new location), then repeat this cycle. Backlight sleep time is set at 2 minutes. The 1.0 cfm unit, sample time is set at one minute, print record, a 1-minute hold time (simulating continuous sampling mode), then repeat this cycle. Backlight would remain on at all times.
| |